Background. Arrhythmogenic cardiomyopathy (ACM) is an inherited disease characterized by fibro-fatty replacement of the myocardium and associated with a risk of sudden cardiac death. The transvenous implantable cardioverter-defibrillator (TV-ICD) is an established therapy, but over time it exposes these patients to lead-related complications. To overcome these limitations, growing evidence supports the subcutaneous ICD (S-ICD) as a valid alternative in patients who do not require bradycardia pacing or cardiac resynchronization. However, some patients with ACM are not candidates for an S-ICD because of pre-implant screening failure or the intrinsic characteristics of the device. More recently, the extravascular ICD (EV-ICD) has been developed: it uses a substernal lead and can deliver antitachycardia pacing (ATP). Purpose. To assess the feasibility and short-term outcomes of the EV-ICD in patients with ACM who are not candidates for, or less suited to, S-ICD implantation. Methods. From July 2024 to September 2026, we evaluated 23 consecutive patients with ACM and an ICD indication, with a non-transvenous system as the preferred strategy. All patients underwent pre-implant S-ICD screening. Twelve patients received an S-ICD and 11 an EV-ICD. Among the latter, the EV-ICD was chosen because of failed S-ICD screening in 8 cases, nickel allergy in 1, low body mass index in 1, and recurrent monomorphic non-sustained ventricular tachycardia for which ATP was desirable in 1. Implantation was planned with pre-procedural chest CT and performed under general anesthesia with systematic defibrillation testing. Follow-up was performed through outpatient visits and continuous remote monitoring. Results. EV-ICD patients had a mean age of 31.8 ± 8.9 years, and 6 (54.5%) were female. The phenotype was left-dominant in 7 (63.6%), and all patients received the EV-ICD for primary prevention. The lead was successfully placed at the first attempt in all patients, with no major intraprocedural complications. Mean R-wave amplitude at implantation was 3.19 ± 1.29 mV (3.14 ± 1.29 mV in patients with failed screening). Defibrillation was successful at 30 J in 9 patients and at 40 J in the remaining 2. During a mean follow-up of 12 ± 8 months, there were no appropriate or inappropriate shocks and no lead dislodgements. One patient required pocket revision for hypersensitivity eczema. One episode of myopotential oversensing was recognized without inappropriate therapy, and no P-wave oversensing was observed. Mean R-wave amplitude in the whole cohort at the latest follow-up was 2.64 ± 0.96 mV. Conclusions. In this single-centre experience, the EV-ICD was feasible and safe in patients with ACM unsuitable for an S-ICD, with effective defibrillation and stable sensing during short-term follow-up. The EV-ICD may therefore represent a valid non-transvenous alternative in young patients with ACM without a pacing indication. Studies with larger cohorts and longer follow-up are needed.

Presupposti dello studio. La cardiomiopatia aritmogena (ACM) è una malattia caratterizzata dalla sostituzione fibroadiposa del miocardio e associata al rischio di morte improvvisa. Il defibrillatore impiantabile transvenoso (TV-ICD) rappresenta una terapia consolidata, ma espone questi pazienti, nel corso del tempo, alle complicanze legate agli elettrocateteri. Per superare questi limiti, si sono accumulate evidenze che indicano il defibrillatore sottocutaneo (S-ICD) come una valida alternativa nei pazienti che non necessitano di stimolazione antibradicardica o di resincronizzazione. Tuttavia, una parte dei pazienti con ACM non è candidabile all'S-ICD per il fallimento dello screening pre-impianto o per le caratteristiche intrinseche del dispositivo. Recentemente è stato sviluppato il defibrillatore extravascolare (EV-ICD), caratterizzato da un elettrocatetere retrosternale e in grado di erogare la stimolazione antitachicardica (ATP). Scopo dello studio. Valutare la fattibilità e gli outcome a breve termine dell'EV-ICD in pazienti con ACM non candidabili o meno adatti all'impianto di S-ICD. Materiali e metodi. Da luglio 2024 a settembre 2026 abbiamo valutato 23 pazienti consecutivi con ACM e indicazione a ICD, considerando come prima scelta un sistema non transvenoso. Tutti i pazienti sono stati sottoposti a screening pre-impianto per S-ICD. Dodici pazienti hanno ricevuto un S-ICD e 11 un EV-ICD. In questi ultimi, l'EV-ICD è stato scelto per fallimento dello screening per S-ICD in 8 casi, allergia al nichel in 1, basso indice di massa corporea in 1 e tachicardia ventricolare non sostenuta monomorfa ricorrente con ATP desiderabile in 1. L'impianto è stato pianificato con TC del torace pre-procedurale ed eseguito in anestesia generale con test di defibrillazione sistematico. Il follow-up è stato condotto ambulatorialmente e con monitoraggio remoto continuo. Risultati. I pazienti con EV-ICD avevano un'età media di 31.8 ± 8.9 anni e 6 (54.5%) erano donne. Il fenotipo era sinistro-dominante in 7 (63.6%), e tutti hanno ricevuto l'EV-ICD in prevenzione primaria. L'elettrocatetere è stato posizionato con successo al primo tentativo in tutti i pazienti, senza complicanze intraprocedurali maggiori. L'onda R media all'impianto era di 3.19 ± 1.29 mV (3.14 ± 1.29 mV nei pazienti con screening fallito). La defibrillazione è risultata efficace a 30 J in 9 pazienti e a 40 J nei restanti 2. A un follow-up medio di 12 ± 8 mesi non si sono verificati shock, appropriati o inappropriati, né dislocazioni dell'elettrocatetere. Un paziente ha richiesto una revisione della tasca per un eczema da ipersensibilità. Un episodio di oversensing da miopotenziali è stato riconosciuto senza terapie inappropriate, e non si è osservato alcun oversensing dell'onda P. L'onda R media dell'intera coorte all'ultimo controllo era di 2.64 ± 0.96 mV. Conclusioni. In questa esperienza monocentrica l'EV-ICD si è dimostrato fattibile e sicuro nei pazienti con ACM non idonei all'S-ICD, con defibrillazione efficace e sensing stabile in un follow-up a breve termine. L'EV-ICD può quindi rappresentare una valida alternativa non transvenosa nei pazienti giovani con ACM senza indicazione al pacing. Sono necessari studi con casistiche più ampie e follow-up più lungo.

ICD extravascolare in pazienti con cardiomiopatia aritmogena non suscettibili di impianto di ICD sottocutaneo: un'esperienza monocentrica

WINNICKI, GIULIA
2024/2025

Abstract

Background. Arrhythmogenic cardiomyopathy (ACM) is an inherited disease characterized by fibro-fatty replacement of the myocardium and associated with a risk of sudden cardiac death. The transvenous implantable cardioverter-defibrillator (TV-ICD) is an established therapy, but over time it exposes these patients to lead-related complications. To overcome these limitations, growing evidence supports the subcutaneous ICD (S-ICD) as a valid alternative in patients who do not require bradycardia pacing or cardiac resynchronization. However, some patients with ACM are not candidates for an S-ICD because of pre-implant screening failure or the intrinsic characteristics of the device. More recently, the extravascular ICD (EV-ICD) has been developed: it uses a substernal lead and can deliver antitachycardia pacing (ATP). Purpose. To assess the feasibility and short-term outcomes of the EV-ICD in patients with ACM who are not candidates for, or less suited to, S-ICD implantation. Methods. From July 2024 to September 2026, we evaluated 23 consecutive patients with ACM and an ICD indication, with a non-transvenous system as the preferred strategy. All patients underwent pre-implant S-ICD screening. Twelve patients received an S-ICD and 11 an EV-ICD. Among the latter, the EV-ICD was chosen because of failed S-ICD screening in 8 cases, nickel allergy in 1, low body mass index in 1, and recurrent monomorphic non-sustained ventricular tachycardia for which ATP was desirable in 1. Implantation was planned with pre-procedural chest CT and performed under general anesthesia with systematic defibrillation testing. Follow-up was performed through outpatient visits and continuous remote monitoring. Results. EV-ICD patients had a mean age of 31.8 ± 8.9 years, and 6 (54.5%) were female. The phenotype was left-dominant in 7 (63.6%), and all patients received the EV-ICD for primary prevention. The lead was successfully placed at the first attempt in all patients, with no major intraprocedural complications. Mean R-wave amplitude at implantation was 3.19 ± 1.29 mV (3.14 ± 1.29 mV in patients with failed screening). Defibrillation was successful at 30 J in 9 patients and at 40 J in the remaining 2. During a mean follow-up of 12 ± 8 months, there were no appropriate or inappropriate shocks and no lead dislodgements. One patient required pocket revision for hypersensitivity eczema. One episode of myopotential oversensing was recognized without inappropriate therapy, and no P-wave oversensing was observed. Mean R-wave amplitude in the whole cohort at the latest follow-up was 2.64 ± 0.96 mV. Conclusions. In this single-centre experience, the EV-ICD was feasible and safe in patients with ACM unsuitable for an S-ICD, with effective defibrillation and stable sensing during short-term follow-up. The EV-ICD may therefore represent a valid non-transvenous alternative in young patients with ACM without a pacing indication. Studies with larger cohorts and longer follow-up are needed.
2024
Extravascular implantable cardioverter defibrillator in arrhythmogenic cardiomyopathy with limitations to subcutaneous ICD: a single center experience
Presupposti dello studio. La cardiomiopatia aritmogena (ACM) è una malattia caratterizzata dalla sostituzione fibroadiposa del miocardio e associata al rischio di morte improvvisa. Il defibrillatore impiantabile transvenoso (TV-ICD) rappresenta una terapia consolidata, ma espone questi pazienti, nel corso del tempo, alle complicanze legate agli elettrocateteri. Per superare questi limiti, si sono accumulate evidenze che indicano il defibrillatore sottocutaneo (S-ICD) come una valida alternativa nei pazienti che non necessitano di stimolazione antibradicardica o di resincronizzazione. Tuttavia, una parte dei pazienti con ACM non è candidabile all'S-ICD per il fallimento dello screening pre-impianto o per le caratteristiche intrinseche del dispositivo. Recentemente è stato sviluppato il defibrillatore extravascolare (EV-ICD), caratterizzato da un elettrocatetere retrosternale e in grado di erogare la stimolazione antitachicardica (ATP). Scopo dello studio. Valutare la fattibilità e gli outcome a breve termine dell'EV-ICD in pazienti con ACM non candidabili o meno adatti all'impianto di S-ICD. Materiali e metodi. Da luglio 2024 a settembre 2026 abbiamo valutato 23 pazienti consecutivi con ACM e indicazione a ICD, considerando come prima scelta un sistema non transvenoso. Tutti i pazienti sono stati sottoposti a screening pre-impianto per S-ICD. Dodici pazienti hanno ricevuto un S-ICD e 11 un EV-ICD. In questi ultimi, l'EV-ICD è stato scelto per fallimento dello screening per S-ICD in 8 casi, allergia al nichel in 1, basso indice di massa corporea in 1 e tachicardia ventricolare non sostenuta monomorfa ricorrente con ATP desiderabile in 1. L'impianto è stato pianificato con TC del torace pre-procedurale ed eseguito in anestesia generale con test di defibrillazione sistematico. Il follow-up è stato condotto ambulatorialmente e con monitoraggio remoto continuo. Risultati. I pazienti con EV-ICD avevano un'età media di 31.8 ± 8.9 anni e 6 (54.5%) erano donne. Il fenotipo era sinistro-dominante in 7 (63.6%), e tutti hanno ricevuto l'EV-ICD in prevenzione primaria. L'elettrocatetere è stato posizionato con successo al primo tentativo in tutti i pazienti, senza complicanze intraprocedurali maggiori. L'onda R media all'impianto era di 3.19 ± 1.29 mV (3.14 ± 1.29 mV nei pazienti con screening fallito). La defibrillazione è risultata efficace a 30 J in 9 pazienti e a 40 J nei restanti 2. A un follow-up medio di 12 ± 8 mesi non si sono verificati shock, appropriati o inappropriati, né dislocazioni dell'elettrocatetere. Un paziente ha richiesto una revisione della tasca per un eczema da ipersensibilità. Un episodio di oversensing da miopotenziali è stato riconosciuto senza terapie inappropriate, e non si è osservato alcun oversensing dell'onda P. L'onda R media dell'intera coorte all'ultimo controllo era di 2.64 ± 0.96 mV. Conclusioni. In questa esperienza monocentrica l'EV-ICD si è dimostrato fattibile e sicuro nei pazienti con ACM non idonei all'S-ICD, con defibrillazione efficace e sensing stabile in un follow-up a breve termine. L'EV-ICD può quindi rappresentare una valida alternativa non transvenosa nei pazienti giovani con ACM senza indicazione al pacing. Sono necessari studi con casistiche più ampie e follow-up più lungo.
ACM
EV-ICD
S-ICD
File in questo prodotto:
File Dimensione Formato  
Tesi specializzazione Winnicki.pdf

Accesso riservato

Dimensione 3.99 MB
Formato Adobe PDF
3.99 MB Adobe PDF

The text of this website © Università degli studi di Padova. Full Text are published under a non-exclusive license. Metadata are under a CC0 License

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.12608/120700