This bachelor’s thesis analyzes the characterization of sensors for continuous glucose monitoring (CGM), framing them within the framework of linear time-invariant (LTI) systems. The work addresses the issue of physiological delay (lag-time) between the blood and interstitial compartments, modeling mass transport through a first-order differential equation based on Fick's law. The analysis in the complex frequency domain, carried out using the Laplace transform, allows defining the system's transfer function and identifying the natural low-pass behavior of the blood-interstitial barrier, characterized by a specific physiological time constant. The model analysis shows that the system is BIBO stable, while the ideal inverse operator is not BIBO stable, and deconvolution exhibits significant ill-conditioning.In particular, the gain of the inverse operator increases with frequency, causing a strong amplification of high-frequency noise components. The research concludes that pure deconvolution is not feasible in real clinical settings without integrating regularization techniques and stochastic filtering, such as the Kalman filter. These methods are essential to balance the reduction of dynamic delay with the stability and reliability of the estimate, helping to ensure the patient’s therapeutic safety.

Questa tesi di laurea triennale analizza la caratterizzazione dei sensori per il monitoraggio continuo del glucosio (CGM) inquadrandoli nel framework dei sistemi lineari tempo-invarianti (LTI). Il lavoro affronta il problema del ritardo fisiologico (lag-time) tra il compartimento ematico e quello interstiziale, modellizzando il trasporto di massa attraverso un’equazione differenziale del primo ordine basata sulla legge di Fick. L’analisi nel dominio della frequenza complessa, condotta tramite la trasformata di Laplace, permette di definire la funzione di trasferimento del sistema e di identificare il comportamento passa-basso naturale della barriera emato-interstiziale, caratterizzato da una specifica costante di tempo fisiologica. L’analisi del modello mostra che il sistema è BIBO stabile, mentre l’operatore inverso ideale non è BIBO stabile e la deconvoluzione presenta un marcato mal-condizionamento. In particolare, il guadagno dell’operatore inverso cresce con la frequenza, determinando una forte amplificazione delle componenti di rumore ad alta frequenza. La ricerca conclude che la deconvoluzione pura non è realizzabile in contesti clinici reali senza l’integrazione di tecniche di regolarizzazione e di filtraggio stocastico, come il filtro di Kalman. Tali metodologie risultano essenziali per bilanciare la riduzione del ritardo dinamico con la stabilità e l’affidabilità della stima, contribuendo a garantire la sicurezza terapeutica del paziente.

Caratterizzazione dei sensori per il monitoraggio continuo del glucosio come sistemi lineari tempo-invarianti

ZOGNO, TOMMASO
2025/2026

Abstract

This bachelor’s thesis analyzes the characterization of sensors for continuous glucose monitoring (CGM), framing them within the framework of linear time-invariant (LTI) systems. The work addresses the issue of physiological delay (lag-time) between the blood and interstitial compartments, modeling mass transport through a first-order differential equation based on Fick's law. The analysis in the complex frequency domain, carried out using the Laplace transform, allows defining the system's transfer function and identifying the natural low-pass behavior of the blood-interstitial barrier, characterized by a specific physiological time constant. The model analysis shows that the system is BIBO stable, while the ideal inverse operator is not BIBO stable, and deconvolution exhibits significant ill-conditioning.In particular, the gain of the inverse operator increases with frequency, causing a strong amplification of high-frequency noise components. The research concludes that pure deconvolution is not feasible in real clinical settings without integrating regularization techniques and stochastic filtering, such as the Kalman filter. These methods are essential to balance the reduction of dynamic delay with the stability and reliability of the estimate, helping to ensure the patient’s therapeutic safety.
2025
Characterization of sensors for continuous glucose monitoring as linear time-invariant systems
Questa tesi di laurea triennale analizza la caratterizzazione dei sensori per il monitoraggio continuo del glucosio (CGM) inquadrandoli nel framework dei sistemi lineari tempo-invarianti (LTI). Il lavoro affronta il problema del ritardo fisiologico (lag-time) tra il compartimento ematico e quello interstiziale, modellizzando il trasporto di massa attraverso un’equazione differenziale del primo ordine basata sulla legge di Fick. L’analisi nel dominio della frequenza complessa, condotta tramite la trasformata di Laplace, permette di definire la funzione di trasferimento del sistema e di identificare il comportamento passa-basso naturale della barriera emato-interstiziale, caratterizzato da una specifica costante di tempo fisiologica. L’analisi del modello mostra che il sistema è BIBO stabile, mentre l’operatore inverso ideale non è BIBO stabile e la deconvoluzione presenta un marcato mal-condizionamento. In particolare, il guadagno dell’operatore inverso cresce con la frequenza, determinando una forte amplificazione delle componenti di rumore ad alta frequenza. La ricerca conclude che la deconvoluzione pura non è realizzabile in contesti clinici reali senza l’integrazione di tecniche di regolarizzazione e di filtraggio stocastico, come il filtro di Kalman. Tali metodologie risultano essenziali per bilanciare la riduzione del ritardo dinamico con la stabilità e l’affidabilità della stima, contribuendo a garantire la sicurezza terapeutica del paziente.
Sistemi LTI
glucosio
deconvoluzione
File in questo prodotto:
File Dimensione Formato  
ZOGNO_TOMMASO.pdf

accesso aperto

Dimensione 658.72 kB
Formato Adobe PDF
658.72 kB Adobe PDF Visualizza/Apri

The text of this website © Università degli studi di Padova. Full Text are published under a non-exclusive license. Metadata are under a CC0 License

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.12608/120041