Background: The number of dental implants being placed worldwide is constantly increasing which consequently leads to a higher probability of adverse events such as the lesion of vital structures during the implant site preparation. In this regard, the utilization of drill stops can be a valuable aid in reducing the risk of undesirable incidents. The purpose of this study is to evaluate the difference in implant site preparation with and without the use of drill stops. Material and methods: 40 implant sites were prepared on pork ribs. The parameters included whether to use stop, operator experience, and the use of irrigation. Subsequently the bone was sectioned, and the depth of the preparation was measured using software. Results: The preparation with the use of stops reached an average depth of 8.01 mm, while without use of stops, it averaged 8.45mm. The difference between the means of the two groups was therefore 0.44mm (P=0.046) and was statistically significant. The average difference in preparation between the experienced and inexperienced operator was 0.47mm (p=0.030). The average difference in preparation with and without irrigation was 0.19mm (P=0.408). Conclusions: The use of a drill stop is associated with a significant improvement in the accuracy of implant site preparation, proving to be a valuable aid, especially for less experienced clinicians. Keywords: Drill Stop, complications, implants, irrigation

Introduzione: Il numero di impianti dentali inseriti nel mondo sta costantemente crescendo con un conseguente aumento delle probabilità di eventi avversi come lesioni delle strutture nobili durante la preparazione del sito implantare. A questo proposito l'impiego di stop per le frese può rappresentare un valido ausilio nella mitigazione del potenziale rischio di eventi indesiderati. Lo scopo principale del presente lavoro è quello di valutare la differenza di preparazione del sito implantare con e senza l’utilizzo di stop per le frese. Materiali e metodi: Sono stati preparati 40 siti implantari su costine di suino. I parametri considerati comprendevano l’utilizzo o meno di stop per la fresa, l’esperienza dell’operatore e l’utilizzo o meno dell’irrigazione. Successivamente l’osso è stato sezionato e la profondità della preparazione è stata misurata con l’ausilio di un software dedicato. Risultati: La preparazione con utilizzo di stop ha raggiunto una profondità media di 8.01 mm, mentre, senza l’utilizzo di stop una media di 8.45mm. La differenza tra le medie dei due gruppi è quindi stata di 0.44 mm (p=0.046) ed è risultata statisticamente significativa. La differenza media di preparazione dell’operatore esperto rispetto al non esperto è stata di 0.47mm(p=0.030). La differenza media di preparazione con e senza acqua è stata di 0.19mm(p=0.408). Conclusioni: L’impiego dello stop per la fresa è associato a un miglioramento significativo nella precisione della preparazione del sito implantare, risultando un valido aiuto soprattutto per il clinico meno esperto. Parole chiave: Drill Stop, complicanze, impianti, l’irrigazione

PREPARAZIONE DEL SITO IMPLANTARE CON E SENZA UTILIZZO DI STOP PER LE FRESE. UNO STUDIO IN VITRO

KHERAD ZARE, ELNAZ
2022/2023

Abstract

Background: The number of dental implants being placed worldwide is constantly increasing which consequently leads to a higher probability of adverse events such as the lesion of vital structures during the implant site preparation. In this regard, the utilization of drill stops can be a valuable aid in reducing the risk of undesirable incidents. The purpose of this study is to evaluate the difference in implant site preparation with and without the use of drill stops. Material and methods: 40 implant sites were prepared on pork ribs. The parameters included whether to use stop, operator experience, and the use of irrigation. Subsequently the bone was sectioned, and the depth of the preparation was measured using software. Results: The preparation with the use of stops reached an average depth of 8.01 mm, while without use of stops, it averaged 8.45mm. The difference between the means of the two groups was therefore 0.44mm (P=0.046) and was statistically significant. The average difference in preparation between the experienced and inexperienced operator was 0.47mm (p=0.030). The average difference in preparation with and without irrigation was 0.19mm (P=0.408). Conclusions: The use of a drill stop is associated with a significant improvement in the accuracy of implant site preparation, proving to be a valuable aid, especially for less experienced clinicians. Keywords: Drill Stop, complications, implants, irrigation
2022
IMPLANT SITE PREPARATION WITH AND WITHOUT DRILL STOPS. AN IN VITRO STUDY
Introduzione: Il numero di impianti dentali inseriti nel mondo sta costantemente crescendo con un conseguente aumento delle probabilità di eventi avversi come lesioni delle strutture nobili durante la preparazione del sito implantare. A questo proposito l'impiego di stop per le frese può rappresentare un valido ausilio nella mitigazione del potenziale rischio di eventi indesiderati. Lo scopo principale del presente lavoro è quello di valutare la differenza di preparazione del sito implantare con e senza l’utilizzo di stop per le frese. Materiali e metodi: Sono stati preparati 40 siti implantari su costine di suino. I parametri considerati comprendevano l’utilizzo o meno di stop per la fresa, l’esperienza dell’operatore e l’utilizzo o meno dell’irrigazione. Successivamente l’osso è stato sezionato e la profondità della preparazione è stata misurata con l’ausilio di un software dedicato. Risultati: La preparazione con utilizzo di stop ha raggiunto una profondità media di 8.01 mm, mentre, senza l’utilizzo di stop una media di 8.45mm. La differenza tra le medie dei due gruppi è quindi stata di 0.44 mm (p=0.046) ed è risultata statisticamente significativa. La differenza media di preparazione dell’operatore esperto rispetto al non esperto è stata di 0.47mm(p=0.030). La differenza media di preparazione con e senza acqua è stata di 0.19mm(p=0.408). Conclusioni: L’impiego dello stop per la fresa è associato a un miglioramento significativo nella precisione della preparazione del sito implantare, risultando un valido aiuto soprattutto per il clinico meno esperto. Parole chiave: Drill Stop, complicanze, impianti, l’irrigazione
impianti
drill stop
complicanze
File in questo prodotto:
File Dimensione Formato  
TESI FINALE PDF.pdf

accesso riservato

Dimensione 1.32 MB
Formato Adobe PDF
1.32 MB Adobe PDF

The text of this website © Università degli studi di Padova. Full Text are published under a non-exclusive license. Metadata are under a CC0 License

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.12608/56687